EPS 45-80 HIGH VOLTAGE GENERATOR SPECIFICATION
RDA Power(*) 30 kW 45 kW 50kW 65 kW 65 kW
RDA kV 40-150        
RDA mA 10-400 10-500 10-630 10-800 10-1000
RDA mAs 0.4-500 0.4-1000 0.4-1000 0.4-1000 0.4-1000
RDA time (s 0.001-10.0       0.001-10( )
Source of AEC Control Ion chamber,photodiode,PMT tube or external DC voltage level
*Other power configurations available 63 s with High Performance Radiography option
FLUORO Continuous Pulsed (0.25-30p/s)
FLUOROkV 40-125 40-125
FLUOROmA 0.2-12.5 16-80
FLUOROmAs -- 0.08-5.0
Source if ABS Control Photodiode,PMT tube or external DC voltage level

Power line
  • 3-Phase line, delta or Wye configuration,
  • 400VAC - 480VAC (50-60Hz)
  • Input lune Voltage+- 10% (automatic adjustment)
Dimensions
Without decorative cabinet With decorative cabinet
441 mm X 578 mmX492 mm (17-4" X 22.7" X 19.4")64- 472 mm X 768 mm X 812 mm (18.64" X 30.2"
75Kg (140-165lb)* *depending on configuration X 32") 79-90Kg (178-198 lb)* *depending on configuration
Options
  • High Performance radiography (12,500mAs and 63-s irradiation time)
  • Auxiliary power supplies 24 VDC or 24 VDC/AC
  • Integrated dual speed starter
  • solid state ion chamber interface
  • Photodiod ion chamber fluoro board
  • Expension boards (Relay, I/O.., etc)
  • High power fluoro (140kV @ 20mA)
  • High performance pulse fluoroscopy
  • Two-tube option (eith cabinet)
  • DPA interface
  • Decorative cabinet
  • hand Switch
  • Custom pedestal stand