| Generator |
| Control |
Microprocessor based with Tactile Membrane Switches |
| Radiographic Mode |
| kVp output |
35 to 100 KVp ( 5 KVp step) |
| mA output |
20 mA Fixed |
| mAs range |
4 to 120 mAs |
| Fluoroscopy Mode |
| KVp output |
30 to 100 kVp ( Motorised 1 KVp Step ) |
| mA output |
0.1 - 3 mA ( Motorised stepless ) |
| Fluoro Timer |
0 to 5 min. cumulative timer buzzer |
| Safety Precaution |
- Temp. Sensor to cut exposure after over heating of x-ray tube head
- Exhaust fan on the top of X-Ray tube head for cooling when you needed long & complicated surgery
|
| X-ray Tube |
Focal spot 0.6 / 1.8 mm |
| TV Camera |
| Type |
High Sensitivity, Low noise Sony CCD camera |
| Image Intensifire |
: |
6" Single / Duel field or 9" Triple field, high gain, low dose image intensifire tube |
| Monitor |
: |
Two no's 17" Color monitors with high density & low noise (Samsung or LG) |
| Memory |
: |
PC Compactable 32 frames memory with full function remote control. Real time zoom with vertical & horizontal panning |
| Optional |
: |
IMS (Image Management System) High frequency generator is also available |
| C - Arm Stand |
| C Orbital movement |
: |
115° ( 90° + 25° ) |
| C - Arm Rotation |
: |
180° each side |
| Motorised vertical movement |
: |
450 mm |
| Forward Movement |
: |
200 mm |
| Right to Left wigwag movement |
: |
+ / - 12.5° |
|
| Indexing For C - Arm Rotaion to get exact AP / LAT Views ( 90° + 90° + 90° + 90° ) |
| Power Requirements : 220 Vac +/- 10%/50 Hz. Line current 18 amps @ 220 vac |
| C-Arm Image Intensiffer Features |
| Mechanical C- Stand |
- Enhanced esthetic compact model can be easily used in small operation theaters also.
- Bigger castors made of hard polymer having 360˚ rotation give very comfortable mobility & maneuverability without steering.
- Silent up-down, motorized movements controlled by 4 limit switches. Effortless mechanical movements with firm & easy locking. All dimensions are maintained with international standard
- Design of indexing gives perfect vertical & horizontal position of C-Arm. The scale need not be referred for there four position at 90˚.
- Fully powder coated unit gives scratch proof / corrosion proof / fumigation proof, international quality powder coating.
- The lightweight Aluminum Alloy reverse bent 'C', gives perfect balance of I. I. Tube & x-ray tube head
|
| Because of its reverse bent design total 'C' can be draped. This ensures safe & secure surgery due to ease of sterilization |
| SPECIAL FEATURES |
- International Standard Aluminum reverse 'C' provides sterilized draping during surgery.
- Large hard polymer wheels to steer instantly with single locking system.
- I. I. Make Thallus, France.
- Sony CCD camera, Japan.
- Microprocessor based digital control panel (No card based machine).
- Gas spring movement ups and downs of C-Arm (Noise free. Less power consumption).
- Indexing for 'C-Arm' rotation to get exact AT/LAT views (90+90+90+90 degrees).
- Unlimited image memory in PC through image management software.
- Print out facility available of stored images in PC.
- Original Intel PIV computer in built.
|
Return to Top |